Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Triplett 4404: (1 year) CAL VER /5520 DATE: 11-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 39 NUMBER OF LINES: 160 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P F V 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Warm-up time: 30 min. - REQUIRED 1.004 DISP Temperature: 20 - 30 degC 1.004 DISP Humidity: <80% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD {DC VOLTAGE} 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT VOLT/OHM 1.007 DISP [32] 5520A NORMAL LO to UUT COM 1.008 HEAD DC VOLTAGE: {400 mV Range} 1.009 5520 400 300.0mV 0.5% 0.1U 2W 2.001 HEAD DC VOLTAGE: {4 V Range} 2.002 5520 4 3.000V 0.5% 0.001U 2W 3.001 HEAD DC VOLTAGE: {40 V Range} 3.002 5520 40 30.00V 0.5% 0.01U 2W 4.001 HEAD DC VOLTAGE: {400 V Range} 4.002 5520 400 300.0V 0.5% 0.1U 2W 5.001 HEAD DC VOLTAGE: {1000 V Range} 5.002 5520 1000 900V 0.5% 2U 2W 6.001 HEAD {AC VOLTAGE} 6.002 HEAD AC VOLTAGE: {400 mV Range} 6.003 5520 400 300.0mV 1% 0.4U 100H SI 2W 7.001 HEAD AC VOLTAGE: {4 V Range} 7.002 5520 4 3.000V 1% 0.003U 100H SI 2W 8.001 HEAD AC VOLTAGE: {40 V Range} 8.002 5520 40 30.00V 1% 0.03U 100H SI 2W 9.001 HEAD AC VOLTAGE: {400 V Range} 9.002 5520 400 300.0V 1% 0.3U 100H SI 2W 10.001 HEAD AC VOLTAGE: {750 V Range} 10.002 5520 750 700V 1.2% 4U 100H SI 2W 11.001 HEAD {FREQUENCY} 11.002 DISP Set UUT to Frequency mode. 11.003 ASK- R 11.004 HEAD FREQUENCY: {100 Hz Range} 11.005 5520 100 90.00H 0.5% 0.01U 1Vpp SI 2W 12.001 HEAD FREQUENCY: {1 kHz Range} 12.002 5520 1000 900.0H 0.5% 0.1U 1Vpp SI 2W 13.001 HEAD FREQUENCY: {10 kHz Range} 13.002 5520 10 9.000kH 0.5% 0.001U 1Vpp SI 2W 14.001 HEAD FREQUENCY: {100 kHz Range} 14.002 5520 100 90.00kH 0.5% 0.01U 1Vpp SI 2W 15.001 HEAD FREQUENCY: {1 MHz Range} 15.002 5520 1000 900.0kH 0.5% 0.1U 1Vpp SI 2W 16.001 ASK+ R 16.002 HEAD {DIODE} 16.003 DISP Set UUT to Diode mode. 16.004 5520 0V S 2W 16.005 EVAL [N]Does {display indicate an overload condition}? 17.001 HEAD {RESISTANCE} 17.002 HEAD RESISTANCE: {40 MOhm Range} 17.003 5520 40 39.00MZ 3% 0.02U 2W 18.001 HEAD RESISTANCE: {4 MOhm Range} 18.002 5520 4000 3900kZ 0.8% 1U 2W 19.001 HEAD RESISTANCE: {400 kOhm Range} 19.002 5520 400 390.0kZ 0.8% 0.1U 2W 20.001 HEAD RESISTANCE: {40 kOhm Range} 20.002 DISP Connect the 5520A to the UUT as follows: 20.002 DISP [32] 5520A AUX HI to UUT V/OHM 20.002 DISP [32] 5520A AUX LO to UUT COM 20.002 DISP [32] 5520A NORMAL HI to UUT V/OHM 20.002 DISP [32] 5520A NORMAL LO to UUT COM 20.002 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 20.002 DISP the NORMAL leads to the SENSE leads AT the UUT. 20.003 5520 40 39.00kZ 0.8% 0.01U CW 21.001 HEAD RESISTANCE: {400 Ohm Range} 21.002 5520 400 390.0Z 0.8% 0.2U CW 22.001 HEAD RESISTANCE: {4 kOhm Range} 22.002 5520 4 3.900kZ 0.8% 0.001U CW 23.001 HEAD {CONTINUITY} 23.002 DISP Set UUT to Continuity mode. 23.003 HEAD CONTINUITY: {400 Ohms} 23.004 5520 400.0Z S CW 23.005 EVAL Is {continuity alarm off}? 24.001 HEAD CONTINUITY: {<40 Ohms} 24.002 5520 35.0Z S CW 24.003 EVAL Is {continuity alarm sounding}? 25.001 5520 0.700V 2% 2W 26.001 HEAD {CAPACITANCE} 26.002 DISP Connect 5520A to the UUT as follows: 26.002 DISP [32] 5520A NORMAL HI to UUT Cx + 26.002 DISP [32] 5520A NORMAL LO to UUT Cx - 26.003 HEAD CAPACITANCE: {4 nF Range} 26.004 5520 4 3.000nF 3% 0.01U 2W 27.001 HEAD CAPACITANCE: {40 nF Range} 27.002 5520 40 30.00nF 3% 0.1U 2W 28.001 DISP Connect the 5520A to the UUT as follows: 28.001 DISP [32] 5520A AUX HI to UUT Cx + 28.001 DISP [32] 5520A AUX LO to UUT Cx - 28.001 DISP [32] 5520A NORMAL HI to UUT Cx + 28.001 DISP [32] 5520A NORMAL LO to UUT Cx - 28.001 DISP Connect the SENSE leads DIRECTLY to the UUT. Connect 28.001 DISP the NORMAL leads to the SENSE leads AT the UUT. 28.002 HEAD CAPACITANCE: {400 nF Range} 28.003 5520 400 300.0nF 3% 1U CW 29.001 HEAD CAPACITANCE: {4 uF Range} 29.002 5520 4 3.000uF 3% 0.01U CW 30.001 HEAD CAPACITANCE: {40 uF Range} 30.002 5520 40 30.00uF 3% 0.1U CW 31.001 HEAD {DC CURRENT} 31.002 RNG 4mA 31.003 DISP Connect 5520A to the UUT as follows: 31.003 DISP [32] 5520A AUX HI to UUT mA 31.003 DISP [32] 5520A AUX LO to UUT COM 31.004 HEAD DC CURRENT: {4 mA Range} 31.005 5520 4 3.000mA 0.8% 0.001U 2W 32.001 HEAD DC CURRENT: {40 mA Range} 32.002 5520 40 30.00mA 0.8% 0.01U 2W 33.001 HEAD DC CURRENT: {400 mA Range} 33.002 5520 400 300.0mA 1% 0.1U 2W 34.001 HEAD {AC CURRENT} 34.002 HEAD AC CURRENT: {4 mA Range} 34.003 5520 4 3.000mA 1.2% 0.001U 100H SI 2W 35.001 HEAD AC CURRENT: {40 mA Range} 35.002 5520 40 30.00mA 1.2% 0.01U 100H SI 2W 36.001 HEAD AC CURRENT: {400 mA Range} 36.002 5520 400 300.0mA 1.2% 0.1U 100H SI 2W 37.001 HEAD AC CURRENT: {10 A Range} 37.002 RNG 10A AC 37.003 DISP Connect the 5520A to the UUT as follows: 37.003 DISP [32] 5520A 20A to UUT 10A 37.003 DISP [32] 5520A AUX LO to UUT COM 37.004 5520 10 10.00A 2% 0.001U 100H SI HC 2W 38.001 HEAD {DC CURRENT} 38.002 HEAD DC CURRENT: {10 A Range} 38.003 5520 10 10.00A 2% 0.01U HC 2W 39.001 END